Journal of Atmospheric and Environmental Optics, Volume. 6, Issue 4, 294(2011)

Preliminary Design of Calibration Device of Micro-Thermometer

Le-zhi ZHAO*, Xiao-qin LIU, and Wei LU
Author Affiliations
  • [in Chinese]
  • show less

    Refractive index structure constant is an important atmospheric parameter. There are many methods to measure , one is using micro-thermometer, but the precise controlled micro-thermometer source is difficult to be realized. Therefore, there isnt a standard micro-thermometer source for the calibration of the micro-thermometer until now. Based on the principle of thermodynamics, using adiabatic compression and expansion of gas method, a temperature controlled device is developed. The uncertainty analysis of the device is proved that the device can be used to calibrate micro-thermometer.

    Tools

    Get Citation

    Copy Citation Text

    ZHAO Le-zhi, LIU Xiao-qin, LU Wei. Preliminary Design of Calibration Device of Micro-Thermometer[J]. Journal of Atmospheric and Environmental Optics, 2011, 6(4): 294

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Aug. 25, 2009

    Accepted: --

    Published Online: Jul. 29, 2011

    The Author Email: Le-zhi ZHAO (zhaolele@mail.ustc.edu.cn)

    DOI:10.3969/j.issn.1673-6141.2011.04.008

    Topics