Acta Optica Sinica, Volume. 33, Issue 2, 217001(2013)

Method to Improve Axial Resolution of Spectral Domain Optical Coherence Tomography

Chong Bo* and hu Yongkai
Author Affiliations
  • [in Chinese]
  • show less

    In spectral domain optical coherence tomographic system, group velocity dispersion of broadband light source reduces axial resolution and leads to the decrease of image resolution and contrast. The decline of axial resolution of spectral domain optical coherence tomography (SD-OCT) caused by dispersion is analyzed and a method based on spectrum correction to improve axial resolution of spectral domain optical coherence tomography is proposed. Spectrum coordinate is processed by second-order coefficients fitting. The coefficients are got by three-step threshold iteration. The axial depth is demodulated by the corrected spectrum. Spectral-domain optical coherence tomography experiments of single-layer thin glass and tomato epidermis tissue show that, compared with traditional method, the axial resolution is improved more than 10 μm and resolution of the two-dimensional tomogram is higher. It is generally applicative for different samples.

    Tools

    Get Citation

    Copy Citation Text

    Chong Bo, hu Yongkai. Method to Improve Axial Resolution of Spectral Domain Optical Coherence Tomography[J]. Acta Optica Sinica, 2013, 33(2): 217001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Medical optics and biotechnology

    Received: Aug. 14, 2012

    Accepted: --

    Published Online: Dec. 5, 2012

    The Author Email: Bo Chong (chongbo@nuaa.edu.cn)

    DOI:10.3788/aos201333.0217001

    Topics