Acta Optica Sinica, Volume. 26, Issue 5, 679(2006)

New Technique for Aberration In-Situ Measurement of a Lithographic Projection System

[in Chinese]1,2, [in Chinese]1, [in Chinese]1, and [in Chinese]1,2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    A new technique for measuring the aberrations of a lithographic projection system in situ is proposed, in which the basic theory used to measure the spherical aberration, coma and astigmatism of the projection system with special measurement mark is analyzed in detail, and the method to calculated the image displacement caused by aberration through aligned position coordinate is presented. Through experiments, the aberration parameters, such as spherical aberration, coma and astigmatism of the projection system are obtained with high accuracy by the technique. Taking into account of the influences of process factors, e.g. resist process, on the image displacement induced by aberration, the new technique avoids the dependence of coma in-situ measurement based on wafer exposure on the image parameters, such as defocus and image tilt.

    Tools

    Get Citation

    Copy Citation Text

    [in Chinese], [in Chinese], [in Chinese], [in Chinese]. New Technique for Aberration In-Situ Measurement of a Lithographic Projection System[J]. Acta Optica Sinica, 2006, 26(5): 679

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: May. 23, 2005

    Accepted: --

    Published Online: Jun. 1, 2006

    The Author Email:

    DOI:

    Topics