Optoelectronics Letters, Volume. 17, Issue 11, 641(2021)

Raman spectroscopic detection using a two-dimensional echelle spectrometer

Rui ZHANG*... Wenyi REN, He WANG, Yuanyuan WANG, Zhenkun LIN and Ziqi HAN |Show fewer author(s)
Author Affiliations
  • School of Science, Northwest Agriculture and Forestry University, Yangling 712100, China
  • show less

    In order to meet the high-resolution and wide spectrum range of the backscattering Raman system, this paper designs and builds a Raman test system based on the echelle spectrometer. In the optical splitting system, compared with the ordinary planar grating spectrometer, the use of the echelle improves the resolution of the system without increasing the volume of the system. The use of intensified charge-coupled device (ICCD) in the detection system improves the signal-to-noise ratio (SNR) and the detection limit of weak spectrum. Finally, the Raman system was spectrally calibrated. The broadband backscattering Raman experimental results are given and discussed. The experimental results show that the system has an excellent application prospect for broadband and high-resolution Raman spectrum measurement.

    Tools

    Get Citation

    Copy Citation Text

    ZHANG Rui, REN Wenyi, WANG He, WANG Yuanyuan, LIN Zhenkun, HAN Ziqi. Raman spectroscopic detection using a two-dimensional echelle spectrometer[J]. Optoelectronics Letters, 2021, 17(11): 641

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Apr. 7, 2021

    Accepted: May. 25, 2021

    Published Online: Jan. 10, 2022

    The Author Email: Rui ZHANG (zhangrui_grating@163.com)

    DOI:10.1007/s11801-021-1065-7

    Topics