Spectroscopy and Spectral Analysis, Volume. 33, Issue 7, 1982(2013)

Impact and Improvement of the Mechanical Shutter on Large-Array Filter-Type Multispectral Imaging System

ZHAO Bao-wei1,2、*, XIANGLI Bin1,2, Lv Qun-bo2, ZHANG Gui-feng2, and LIU Jiu2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    The present paper analyzes the impact of mechanical shutter on the spectral image acquisition and processing of large-array filter-type multispectral imaging system. The final image quality relies highly on the mechanical shutter due to the fluctuation at exposure time. The conventional mechanical shutter’s structure and driving method was analyzed to find out the key fact for its poor stability. An improved method of mechanical transmission and circuit driving was proposed. Laboratory experiments showed that with the improved design strategy, the maximum rate of change between adjacent exposures was reduced from 15.05% to 0.96%, which is a great improvement of the exposure time stability. Field test was also carried out and the results show that the combined color images are closer to the realistic targets and no abrupt color change iwas found. It’s of great significance for practical application in multispectral image process, interpretation and target identification.

    Tools

    Get Citation

    Copy Citation Text

    ZHAO Bao-wei, XIANGLI Bin, Lv Qun-bo, ZHANG Gui-feng, LIU Jiu. Impact and Improvement of the Mechanical Shutter on Large-Array Filter-Type Multispectral Imaging System[J]. Spectroscopy and Spectral Analysis, 2013, 33(7): 1982

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Dec. 13, 2012

    Accepted: --

    Published Online: Sep. 30, 2013

    The Author Email: Bao-wei ZHAO (zbw8200980@163.comReferences)

    DOI:10.3964/j.issn.1000-0593(2013)07-1982-05

    Topics