Laser & Optoelectronics Progress, Volume. 56, Issue 10, 101103(2019)

Spectral Calibration Method for Spectral Domain Coherence Tomography with Ultra-Wideband Light Source

Wenjuan Cao, Wanrong Gao*, and Xiupin Wu
Author Affiliations
  • School of Electronic and Optical Engineering, Nanjing University of Science and Technology, Nanjing, Jiangsu 210094, China
  • show less

    Based on spectral domain optical coherence tomography with ultra-wideband light source, one Doppler shift calibration method suitable for a practical system is proposed, in which the mirror motion in the traditional method is replaced by a moving optical-path module, thereby enabling its applications in actual situations. After acquiring a series of time-varying spectra in motion, the proposed method is used to analyze the spectral distribution range via Fourier transform. Compared with the traditional calibration method, the proposed method is more precise and the axial resolution of the system is improved by 0.34 μm. Thus, this method improves the imaging effect in actual biological tissue imaging.

    Tools

    Get Citation

    Copy Citation Text

    Wenjuan Cao, Wanrong Gao, Xiupin Wu. Spectral Calibration Method for Spectral Domain Coherence Tomography with Ultra-Wideband Light Source[J]. Laser & Optoelectronics Progress, 2019, 56(10): 101103

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Oct. 12, 2018

    Accepted: Dec. 20, 2018

    Published Online: Jul. 4, 2019

    The Author Email: Gao Wanrong (wgao@njust.edu)

    DOI:10.3788/LOP56.101103

    Topics