Acta Optica Sinica, Volume. 33, Issue 10, 1028001(2013)
Quantitative Analysis of the Measurement Precision of Phase Retrieval Wavefront Sensor
In order to validate the performance of phase retrieval wavefront sensor (PRWS), an experimental setup based on the PRWS method comparing with Zygo interferometer is built. The liquid crystal spatial light modulator (LC-SLM) is utilized to produce single aberration and random aberration, and the ability of PRWS measurement for any aberration is validated. The experimental results demonstrate that agreement is obtained among the errors distribution, peak valley (PV) value and root mean square (RMS) value of Zygo interferometer. The measurement precision of wavefront is 3λ/1000 (RMS). Therefore, the feasibility and accuracy of the proposed method can be confirmed.
Get Citation
Copy Citation Text
Ma Xinxue, Wang Jianli, Wang Bin, Wang Zongyang. Quantitative Analysis of the Measurement Precision of Phase Retrieval Wavefront Sensor[J]. Acta Optica Sinica, 2013, 33(10): 1028001
Category: Remote Sensing and Sensors
Received: Jan. 25, 2013
Accepted: --
Published Online: Sep. 25, 2013
The Author Email: Xinxue Ma (maxinxue@gmail.com)