Spectroscopy and Spectral Analysis, Volume. 30, Issue 9, 2581(2010)

Study on the Method of Selecting Hadamard Spectrophotometer’s Multi-Slits

FAN Yu1,2、*, WU Yi-hui1, XUAN Ming1, ZHANG Ping1, ZHOU Lian-qun1,2, and LIU Yong-shun1,2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    In order to realize high throughput of Hadamard spectrometer, the performance of four Hadamard S slits are compared. Multi-micro-silicon slits were produced by MEMS technique. Based on diffraction theory, the formula with Hadamard S matrix is presented. The phenomenon of diffraction is analyzed by using the software of Matlab. The comparison was made between the Matlab simulating picture and the actual diffract picture of S slits on throughput. Experiment results indicate the precisions of the system. The influence of the factors such as spot distance, slit dimension and the matrix array on the throughput was analyzed, then the diffracting way of S slit was offered. The results of emulation are consistent with test, and proved that the design method on the slit in the paper is right. The way of analysis can provide some theoretic and application foundations for slit design.

    Tools

    Get Citation

    Copy Citation Text

    FAN Yu, WU Yi-hui, XUAN Ming, ZHANG Ping, ZHOU Lian-qun, LIU Yong-shun. Study on the Method of Selecting Hadamard Spectrophotometer’s Multi-Slits[J]. Spectroscopy and Spectral Analysis, 2010, 30(9): 2581

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Nov. 23, 2009

    Accepted: --

    Published Online: Jan. 26, 2011

    The Author Email: Yu FAN (fanyu126@126.com)

    DOI:

    Topics