Acta Optica Sinica, Volume. 35, Issue 1, 112007(2015)

Study of Anamorphose Correction of Astigmatism-Corrected Czerny-Turner Imaging Spectrometer

Tang Yi*, Zheng Cheng, Nan Yibing, Li Fei, and Guo Lei
Author Affiliations
  • [in Chinese]
  • show less

    Correcting astigmatism by using Czerny- Turner imaging spectrometer can change system focal length, which leads to the difference of magnification in tangential direction and sagittal direction of the image. The common characteristics of astigmatism correction theories of Czerny- Turner imaging spectrometers (using cylindrical mirror and toroidal mirror) are analyzed. And the correction formulas, methods of anamorphose and their common characteristics are given. The validity of correction formula of anamorphose are verified with theoretical calculation, ray tracing simulation, and actual measurement. Meanwhile, this method of image correction is useful to the anamorphose correction research of off- axis structure dispersion imaging spectrometer in other operating conditions of collimation light.

    Tools

    Get Citation

    Copy Citation Text

    Tang Yi, Zheng Cheng, Nan Yibing, Li Fei, Guo Lei. Study of Anamorphose Correction of Astigmatism-Corrected Czerny-Turner Imaging Spectrometer[J]. Acta Optica Sinica, 2015, 35(1): 112007

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Apr. 15, 2014

    Accepted: --

    Published Online: Dec. 15, 2014

    The Author Email: Yi Tang (tangyi4510@bit.edu.cn)

    DOI:10.3788/aos201535.0112007

    Topics