Optical Instruments, Volume. 36, Issue 5, 409(2014)

The application and consideration of haze meter’s measurement and calibration

LIN Yan*... ZHOU Xiaofeng, HU Zifeng and XU Zikai |Show fewer author(s)
Author Affiliations
  • [in Chinese]
  • show less

    The practical problems are introduced which are encountered in the instrument calibration and JJF 1303-2011 ‘Calibration Specification for Haze meter’ in the practical using. The problems mainly focus on three parts:the influence of environmental factors over the measurement results, the indication error of haze standard plate, and the flaw of producing haze meter. Then the improved suggestions are given for the relevant reference.