Precise measurement of gold nanorods by using depolarized dynamic light scattering apparatus

Zhen CUI1,2, Lu HUANG2、*, Miao SUN2, Si-tian GAO2, Jin-hui CAI1, Wei LI2, Yu-shu SHI2, and Qi LI2
Author Affiliations
  • 1China Jiliang University, Hangzhou 310018, China
  • 2National Institute of Metrology (NIM), Beijing 100029, China
  • show less

    A precise and noninvasive method for the size and shape measurement of gold nanorods (GNRs) has been proposed based on depolarized dynamic light scattering (DDLS). A home-made DDLS apparatus has been established. By applying depolarized optical path with precise alignment method, the signal-to-noise ratio (SNR) of this apparatus is highly improved. GNRs with three different diameter and length has been precisely measured by using DDLS method as well as scanning electron microscopy (SEM). The thickness of adsorption layer of cetyltrimethylammonium bromide (CTAB) in solution has been taken into consideration. Results show that size measurement of GNRs by using DDLS method agrees very well with that by using SEM. In addition, it is shown that the extinction spectroscopy strongly limited the application of DDLS method by affecting the effective scattering light intensity. Proper laser wavelength should be chosen before the application of this method.

    Tools

    Get Citation

    Copy Citation Text

    CUI Zhen, HUANG Lu, SUN Miao, GAO Si-tian, CAI Jin-hui, LI Wei, SHI Yu-shu, LI Qi. Precise measurement of gold nanorods by using depolarized dynamic light scattering apparatus[J]. Optoelectronics Letters, 2021, 17(3): 170

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Sep. 9, 2020

    Accepted: Oct. 26, 2020

    Published Online: Sep. 2, 2021

    The Author Email: Lu HUANG (huangl@nim.ac.cn)

    DOI:10.1007/s11801-021-0140-4

    Topics