Acta Optica Sinica, Volume. 30, Issue 4, 1026(2010)

Investigation on Radiation Calibration System of Vacuum Ultraviolet Detector

Wang Rui1,2、*, Wang Shurong1, Li Futian1, Liu Haibo1, Wang Longqi1,2, and Huang Yu1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    A high-accuracy radiation calibration system of vacuum ultraviolet (VUV) detector is established,to meet the development of the ultraviolet (UV) and VUV remote quantification. In 120-220 nm spectral range,the absolute responsivity was transferred to EMR541-09-17 ultraviolet phototube detector,on the basic of national institute of standards and technology (NIST) ultraviolet standard detector. According to the guidelines for evaluating and expressing the uncertainty of NIST measurement,the relative uncertainty of transferring is less than 2.4%. The aim of high accuracy radiation calibration of ultraviolet detector in VUV region has been obtained.

    Tools

    Get Citation

    Copy Citation Text

    Wang Rui, Wang Shurong, Li Futian, Liu Haibo, Wang Longqi, Huang Yu. Investigation on Radiation Calibration System of Vacuum Ultraviolet Detector[J]. Acta Optica Sinica, 2010, 30(4): 1026

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Mar. 23, 2009

    Accepted: --

    Published Online: Apr. 20, 2010

    The Author Email: Rui Wang (darui9999@yahoo.com.cn)

    DOI:10.3788/aos20103004.1026

    Topics