Acta Photonica Sinica, Volume. 44, Issue 3, 311004(2015)

Point Spread Function of Pinhole Imaging System for Deviating from Center of Incident γ-Ray

MA Qing-li1、*, TANG Shi-biao2, and WU Yan-hua1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    The Point Spread Function(PSF) of a pinhole imaging system for γ-Ray has been studied through Monte Carlo method.The PSFs under 0, 0.5, 1, 1.5 and 2 pixel bias five different conditions have been obtained and analyzed by fitting the simulating data with Gauss function, the corresponding Modulation Transfer Function(MTF) can be obtained, and the spatial resolutions under these five conditions are compared.As the results show, when the excursion is little, the PSF′ deviation obtained by Gauss function fitting will be less and can meet the requirement of accuracy, but when the excursion is larger, the fitting deviation will be more. In addition, a ideal baffle between pinhole and detector will reduce the error and improve the spatial resolution remarkably.

    Tools

    Get Citation

    Copy Citation Text

    MA Qing-li, TANG Shi-biao, WU Yan-hua. Point Spread Function of Pinhole Imaging System for Deviating from Center of Incident γ-Ray[J]. Acta Photonica Sinica, 2015, 44(3): 311004

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Oct. 8, 2014

    Accepted: --

    Published Online: Apr. 14, 2015

    The Author Email: Qing-li MA (maql@ustc.edu.cn)

    DOI:10.3788/gzxb20154403.0311004

    Topics