Acta Optica Sinica, Volume. 27, Issue 3, 457(2007)

Optical Radiometric Calibration Based on Correlated Photons

[in Chinese]*, [in Chinese], [in Chinese], [in Chinese], and [in Chinese]
Author Affiliations
  • [in Chinese]
  • show less

    A method of optical radiometric calibration based on correlated photons was proposed to improve calibration precision. It was independent of the methods based on radiometric source or detectors. With 351 nm laser to pump β-Barium Borate(BBO)crystal, a radiometric calibration system based on correlated photons had been set up, and its principle and experimental setup were analyzed. Coincident metrical system was given, and the experimental result is discussed. The results showed that the calibrated quantum efficiency of photomultiplier (R2949) at 702 nm was (5.96±0.003)% , the relative error of experimental result was about 1.0% with respect to the typical quantum efficiency of photomultiplier (R2949)at this wavelength, and relative combined uncertainty of the results was about 2.61%.

    Tools

    Get Citation

    Copy Citation Text

    [in Chinese], [in Chinese], [in Chinese], [in Chinese], [in Chinese]. Optical Radiometric Calibration Based on Correlated Photons[J]. Acta Optica Sinica, 2007, 27(3): 457

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Mar. 29, 2006

    Accepted: --

    Published Online: Mar. 15, 2007

    The Author Email: (yfeng@aiofm.ac.cn)

    DOI:

    Topics