Optics and Precision Engineering, Volume. 17, Issue 5, 980(2009)

Improving accuracy of biochemical analyzer by uniforming integral time data

LI Zheng-gang1,2、*, WU Yi-hui1, XUAN Ming1, and WANG Yi-fan1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    With the aim to improve the performance of a biochemical analyzer,an uniform integral time method for CCD data acquisition was presented in the paper. Different integral time data were calculated based on a Photo Diode Array(PDA) driving program and the photoelectrical transformation principle,and the linearity region of the biochemical analyzer was confirmed according to both performance parameters and experiments. Then,different integral time dark currents were recorded in the experiments and they were removed from an original spectrum to ensure the analyzer error in minimum. The experimental results show that the precision of the analyzer has been improved by 25%,and the work efficiency is 1~3 times higher than those of the original methods. It is proved that the using the proposed method can simplify the experimental procedures and can provide the higher precision for the analyzer even in sample concentration to be changed greatly.

    Tools

    Get Citation

    Copy Citation Text

    LI Zheng-gang, WU Yi-hui, XUAN Ming, WANG Yi-fan. Improving accuracy of biochemical analyzer by uniforming integral time data[J]. Optics and Precision Engineering, 2009, 17(5): 980

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Jul. 23, 2008

    Accepted: --

    Published Online: Oct. 28, 2009

    The Author Email: Zheng-gang LI (leegon@163.com)

    DOI:

    CSTR:32186.14.

    Topics