Optics and Precision Engineering, Volume. 19, Issue 12, 2870(2011)

Spectral calibration of imaging spectrometer with convex grating

QI Xiang-dong1,*... HAN Peng-peng1, PAN Ming-zhong1, and CUI Ji-cheng12 |Show fewer author(s)
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    A spectral calibration facility consisting of a monochromator and a collimator was established to calibrate the Offner imaging spectrometer with a convex grating designed by ourselves. The self-designed software for sampling and processing spectral data was used to calibrate the Offner imaging spectrometer and analyze obtained data. Calibrated results show that the calibration facility has a compact conformation, better commonality and high calibration precision, and its spectral range is better than 400-800 nm and spectral resolution is superior to 10 nm, which can satisfy the calibration need in practice.

    Tools

    Get Citation

    Copy Citation Text

    QI Xiang-dong, HAN Peng-peng, PAN Ming-zhong, CUI Ji-cheng. Spectral calibration of imaging spectrometer with convex grating[J]. Optics and Precision Engineering, 2011, 19(12): 2870

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Apr. 22, 2011

    Accepted: --

    Published Online: Dec. 22, 2011

    The Author Email: Xiang-dong QI (chinagrating@263.net)

    DOI:10.3788/ope.20111912.2870

    Topics