Chinese Journal of Lasers, Volume. 41, Issue 1, 108006(2014)

Layout Optimization of Three-Dimensional Coordinate Measurement System Based on Laser Multi-Lateration

Hu Jinzhong*, Yu Xiaofen, Peng Peng, and Huang Kaihui
Author Affiliations
  • [in Chinese]
  • show less

    The three-dimensional (3D) coordinate measuring system based on laser multi-lateration can realize the non-target self-calibration. Its measurement accuracy is influenced by the self-calibration accuracy. However, the layout of the system is one of main impact factors on the self-calibration accuracy. By theoretical analysis of the non-target self-calibration model, the result of the system layout optimization is the tri-rectangular pyramid layout. Simulation results show that the tri-rectangular pyramid layout can improve the self-calibration accuracy, thus can ensure the system accuracy.

    Tools

    Get Citation

    Copy Citation Text

    Hu Jinzhong, Yu Xiaofen, Peng Peng, Huang Kaihui. Layout Optimization of Three-Dimensional Coordinate Measurement System Based on Laser Multi-Lateration[J]. Chinese Journal of Lasers, 2014, 41(1): 108006

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: measurement and metrology

    Received: Jul. 23, 2013

    Accepted: --

    Published Online: Dec. 24, 2013

    The Author Email: Jinzhong Hu (hujinzhong402@126.com)

    DOI:10.3788/cjl201441.0108006

    Topics