Spectroscopy and Spectral Analysis, Volume. 29, Issue 9, 2375(2009)

Development of a New FTIR Car-Borne Oil Content Analyzer

YANG Kun*, LEI Jun-feng, ZENG Li-bo, and ZHU Qiu-ping
Author Affiliations
  • [in Chinese]
  • show less

    Turlun structure was introduced into the Fourier transform interference system in the present paper in order to decrease the size and cost of the whole system and to increase the ability of vibration resistance to meet the requirement of car-borne outdoor monitoring application. The speed deviation of the voice coil motor (VCM) driving system, which is based on feedback of interference wavelength of reference laser and DSP control, is within 5%. The standard oil sample was used to test the performance of the spectrometer. The result shows that the highest resolution of the FTIR oil detector is 2 cm-1, and the SNR reached 2 000∶1. Compared with old grating spectrometers, this kind of instrument is faster, and the quality of spectrum was increased significantly. The portable feature of this instrument will be more convenient for the water pollution detection in field.

    Tools

    Get Citation

    Copy Citation Text

    YANG Kun, LEI Jun-feng, ZENG Li-bo, ZHU Qiu-ping. Development of a New FTIR Car-Borne Oil Content Analyzer[J]. Spectroscopy and Spectral Analysis, 2009, 29(9): 2375

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Aug. 15, 2008

    Accepted: --

    Published Online: May. 26, 2010

    The Author Email: Kun YANG (quent.young@163.com)