Acta Optica Sinica, Volume. 29, Issue 7, 1956(2009)

Optical Fiber Sensor for Liquid Refractive Index Based on Fiber Optic Taper

Xiong Yikun* and Huang Xuguang
Author Affiliations
  • [in Chinese]
  • show less

    A new refractive index fiber optic sensor with low cost, easy operation and the capability of multipoint parallel measurement is explored. A splitter is used in the refractive index fiber optic sensor to achieve multipoint parallel measurement. A high stable amplified spontaneous emission light source and three optical power meters are used as broadband source and detectors, respectively. A simple model is established, and the relation between the ratio of the light power carried by the core versus the total light power of a fiber and the external refractive index (RI) is analyzed in detail. In the experiment, the transmission losses of sample solution with different refractive indices are measured, the calibration curve between transmission loss (TL) and RI by comparison is obtained. It is found that the higher the RI is, the bigger the TL is. The experiment verifies that it can improve the system sensitivity greatly by introducing a fiber Bragg grating.

    Tools

    Get Citation

    Copy Citation Text

    Xiong Yikun, Huang Xuguang. Optical Fiber Sensor for Liquid Refractive Index Based on Fiber Optic Taper[J]. Acta Optica Sinica, 2009, 29(7): 1956

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Devices

    Received: Aug. 4, 2008

    Accepted: --

    Published Online: Jul. 20, 2009

    The Author Email: Yikun Xiong (xyk117@163.com)

    DOI:

    Topics