  • show less

    We experimentally evaluate and correct the non-equivalence between electrical and radiative heating of solar irradiance absolute radiometer to compensate the systematic error of radiant power measurement at ambient pressure. A relative difference of the order of 0.08%–0.27% between electrical and radiative heating sensitivities is shown, and the resulting non-equivalence correction factor is calculated. The radiant power measurement equation is modified using the non-equivalence correction factor, a systematic deviation of 0.19% of radiant power measurement is hence eliminated.

    Tools

    Get Citation

    Copy Citation Text

    Zhenling Yang, Wei Fang, Yang Luo, Zhiwei Xia. Experimental characterization and correction of non-equivalence of Solar Irradiance Absolute Radiometer[J]. Chinese Optics Letters, 2014, 12(10): 101202