Acta Optica Sinica, Volume. 28, Issue s2, 271(2008)

Evaluation of Uncertainty for CO Monitor Based on Tunable Diode Laser Absorption Spectroscopy

Li Honglian*, Li Xiaoting, Li Jinhai, Han Bing, and Dong Fang
Author Affiliations
  • [in Chinese]
  • show less

    The concentration of CO is measured by tunable diode laser absorption spectroscopy (TDLAS) technology. The experimental data of measurement signals are inversely processed by applying the overall second harmonic least squares data processing algorithm. The principal sources of measurement results uncertainty are analyzed during measurement of CO concentration by TDLAS. The uncertainty components, standard uncertainty, and synthesis uncertainty caused by various factors are evaluated with the direct evaluation method. The results indicate that the major factors affecting measurement uncertainty are the uncertainty of the apparatus, concentration definite value uncertainty of calibration gas, and the uncertainty indicated by ambient temperature variation and supply voltage fluctuation.

    Tools

    Get Citation

    Copy Citation Text

    Li Honglian, Li Xiaoting, Li Jinhai, Han Bing, Dong Fang. Evaluation of Uncertainty for CO Monitor Based on Tunable Diode Laser Absorption Spectroscopy[J]. Acta Optica Sinica, 2008, 28(s2): 271

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: OPTOELECTRONICS

    Received: --

    Accepted: --

    Published Online: Jan. 5, 2009

    The Author Email: Honglian Li (lihonglian@hbu.cn)

    DOI:

    Topics