Chinese Optics Letters, Volume. 8, Issue 4, 407(2010)

Automated damage diagnostic system for laser damage threshold tests

Xiaofeng Liu1,2, Dawei Li1, Yuan’an Zhao1, Xiao Li1,2, Xiulan Ling1,2, and Jianda Shao1
Author Affiliations
  • 1Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Shanghai 201800, China
  • 2Graduate University of Chinese Academy of Sciences, Beijing 100049, China
  • show less

    An automated damage diagnostic system for collecting plasma flash is developed to diagnose damage in a laser-induced damage threshold (LIDT) test system. Experiment is done to verify the accuracy of this system and analyze the relationship between the plasma signals and the damage morphologies. The results obtained by the system are found to be in excellent agreement with those obtained by the much laborious method of Normaski microscope. Results show that plasma signals above 1 V correspond to the damage morphology of surface discolorations with or without pits in their centers, and plasma signals below or just around 1 V correspond to the damage morphology of pits. The misdiagnosis is attributed to contaminations and air breakdown.

    Tools

    Get Citation

    Copy Citation Text

    Xiaofeng Liu, Dawei Li, Yuan’an Zhao, Xiao Li, Xiulan Ling, Jianda Shao. Automated damage diagnostic system for laser damage threshold tests[J]. Chinese Optics Letters, 2010, 8(4): 407

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Aug. 17, 2009

    Accepted: --

    Published Online: Apr. 20, 2010

    The Author Email:

    DOI:10.3788/COL20100804.0407

    Topics