Acta Optica Sinica, Volume. 30, Issue 11, 3082(2010)

An Automatic Defocus Compensation Method for Human Eye Hartmann-Shark Wave-Front Aberrometer

Bao Hua1,2,3、*, Rao Changhui2,3, Zhang Yudong2,3, Dai Yun2,3, Yu Xiang2,3, and Rao Xunjun2,3
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less

    An automatic defocus compensation method for human eye Hartmann-Shark (HS) wave-front aberrometer is proposed. According to the analysis of different geometric features of HS light spots during focalizing process, estimation algorithm is constructed to remove large-scale defocus amount, and then the residual high-order ocular aberrations can be measured by aberrometer successfully. Experimental results proved that the method has good robustness to control noise and can achieve accurate automatic defocus compensation from ocular aberrations, which is very useful method to human eye HS wave-front aberrometer.

    Tools

    Get Citation

    Copy Citation Text

    Bao Hua, Rao Changhui, Zhang Yudong, Dai Yun, Yu Xiang, Rao Xunjun. An Automatic Defocus Compensation Method for Human Eye Hartmann-Shark Wave-Front Aberrometer[J]. Acta Optica Sinica, 2010, 30(11): 3082

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Atmospheric Optics and Oceanic Optics

    Received: Feb. 2, 2010

    Accepted: --

    Published Online: Nov. 16, 2010

    The Author Email: Hua Bao (hbao.ioe@gmail.com)

    DOI:10.3788/aos20103011.3082