Opto-Electronic Engineering, Volume. 42, Issue 3, 44(2015)

Calibrating Laser Plane of Scanning Probe with Two Lasers

ZHOU Aiguo1,*... GE Weiliang1, LU Minxun1, Lü Gang2 and ZHOU Fei1 |Show fewer author(s)
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    A field calibration for laser scanning probe is put forward. Based on the model of camera calibration, the mathematical model of the laser plane is built. In order to get the more accuracy feature points, a new method using two alternate laser stripes which are not parallel to project on the chessboard is designed. The target is put in several different positions within the field of the camera vision aiming to get more feature points. All these feature points are fitted to an ideal plane after coordinate transformation. And the equation of the laser plane is obtained in camera coordinate. Experimental results indicate that the mean calibration error is less than 0.028mm. Using CMM to test the distance precision of the laser plane, the results show that the max distance error is 0.031mm. Experimental results indicate that the calibration technique has high accuracy and this method is suitable for field calibration.

    Tools

    Get Citation

    Copy Citation Text

    ZHOU Aiguo, GE Weiliang, LU Minxun, Lü Gang, ZHOU Fei. Calibrating Laser Plane of Scanning Probe with Two Lasers[J]. Opto-Electronic Engineering, 2015, 42(3): 44

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: May. 7, 2014

    Accepted: --

    Published Online: Mar. 23, 2015

    The Author Email: Aiguo ZHOU (zhouaiguo@tongji.edu.cn)

    DOI:10.3969/j.issn.1003-501x.2015.03.008

    Topics