Spectroscopy and Spectral Analysis, Volume. 31, Issue 2, 521(2011)
[in Chinese]
A wood slice was used as absorber to transfer liquid sample to solid sample in order to solve the problems existing in directly analyzing aqueous solutions with laser-induced breakdown spectroscopy (LIBS). An optical-electrical dual pulse LIBS (OEDP-LIBS) technique was first used to enhance atomic emission of mercury in laser-induced plasma. The calibration curves of mercury were obtained by typical single pulse LIBS and OEDP-LIBS techniques. The limit of detection (LOD) of mercury in these two techniques reaches 2.4 and 0.3 mg·L-1, respectively. Under current experimental conditions, the time-integrated atomic emission of mercury at 253.65 nm was enhanced 50 times and the LOD of mercury was improved by one order, if comparing OEDP-LIBS to single pulse LIBS. The required time for a whole analysis process is less than 5 minutes. As the atomic emission of mercury decays slowly while increasing the delay time between electrical pulse and laser pulse, increasing the electrical pulse width can further enhance the time integrated intensity of mercury emission and improve the detection sensitivity of mercury by OEDP-LIBS technique.
Get Citation
Copy Citation Text
ZHANG Qian, XIONG Wei, CHEN Yu-qi, LI Run-hua. [J]. Spectroscopy and Spectral Analysis, 2011, 31(2): 521
Received: Mar. 1, 2010
Accepted: --
Published Online: Mar. 24, 2011
The Author Email: Qian ZHANG (z.qian06@mail.scut.edu.cn)
CSTR:32186.14.