Optics and Precision Engineering, Volume. 14, Issue 5, 929(2006)

Improvement of spectral domain OCT image quality by linear interpolation and CCD response compensation

ZHANG Tae-Sok1...2,*, LI Gang1, ZHENG Yu1, LIN Ling1 and LIU Jing-jing1 |Show fewer author(s)
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Data collected by CCD camera is a function of wavelength in Spectral-Domain Optical Coherence Tomography(SDOCT) system.By consideration of a wavelength-depended response characteristics of CCD and the Fourier transform relationship between time(distance) and frequency(wave-number),the CCD data must be responsivity-compensated and resampled by linear interpolation in frequency space.Based on the above analysis,the CCD response compensation method and the linear interpolation arithmetic were suggested and tested by the sample imaging.Results show that signal-to-noise ratio(SNR) increases a 30% in proposed method and arithmetic mentioned above,and the OCT image quality is greatly improved.

    Tools

    Get Citation

    Copy Citation Text

    ZHANG Tae-Sok, LI Gang, ZHENG Yu, LIN Ling, LIU Jing-jing. Improvement of spectral domain OCT image quality by linear interpolation and CCD response compensation[J]. Optics and Precision Engineering, 2006, 14(5): 929

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Feb. 14, 2006

    Accepted: --

    Published Online: Feb. 28, 2010

    The Author Email: Tae-Sok ZHANG (zhangtaesok@eyou.com)

    DOI:

    CSTR:32186.14.

    Topics