Acta Optica Sinica, Volume. 32, Issue 4, 412001(2012)

Calibration of Binocular Stereo Measurement System Based on Calibration Cross

Sun Nan*, Zhang Liyan, Ye Nan, and Wang Hongtao
Author Affiliations
  • [in Chinese]
  • show less

    A calibration method of binocular stereo measurement system is proposed based on calibration cross. The intrinsic parameters of the two cameras are calculated precisely and the three-dimensional (3D) target point sets are reconstructed in each camera coordinate system respectively by using the multiple-view geometrical constraint and a generic sparse bundle adjustment. Then the extrinsic parameters are calculated by using the rigid-body transform of the two 3D point sets. This method is simple and flexible for field operation because it only needs the calibration cross to be photographed by two cameras simultaneously once and then separately several times. Based on a binocular stereo measurement system calibrated by the proposed method, a staff gauge of 611.800 mm long is measured many times. The result shows that the average measurement value is 611.776 mm, and standard deviation is 0.030 mm.

    Tools

    Get Citation

    Copy Citation Text

    Sun Nan, Zhang Liyan, Ye Nan, Wang Hongtao. Calibration of Binocular Stereo Measurement System Based on Calibration Cross[J]. Acta Optica Sinica, 2012, 32(4): 412001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Sep. 28, 2011

    Accepted: --

    Published Online: Feb. 20, 2012

    The Author Email: Nan Sun (happy_sunnan@163.com)

    DOI:10.3788/aos201232.0412001