Journal of Atmospheric and Environmental Optics, Volume. 14, Issue 5, 345(2019)

Trace Cr and Cu Analysis in Aluminum Alloy by Double Pulse Laser-Induced Breakdown Spectroscopy

YANGLichao 1、*, Ruizhao YANG2, Xuejiao SU1, Youli YU1, and Weidong ZHOU1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    A collinear dual pulse laser induced breakdown spectroscopy (DP-LIBS) has been used for trace element analysis in aluminum alloy. The dependence of signal intensity on the inter-pulse delay time has been carefully investigated, which gives an optimized delay time of about 8~9 μs. For the two selected analytical lines Cu I 324.75 nm and Cr I 425.43 nm, more than 10 folds enhancement of line intensities have been obtained compared to that of single-pulse laser induced breakdown spectroscopy(SP-LIBS). Calibration curves for quantitative measurement of Cu and Cr were derived using both SP-LIBS and DP-LIBS techniques. The limit of detection (LOD) of elements Cu and Cr are found to be 21.46 and 4.26 μg/g respectively with DP-LIBS, while the LOD of elements Cu and Cr are 169.5 and 94.5 μg/g with SP-LIBS.

    Tools

    Get Citation

    Copy Citation Text

    YANGLichao, YANG Ruizhao, SU Xuejiao, YU Youli, ZHOU Weidong. Trace Cr and Cu Analysis in Aluminum Alloy by Double Pulse Laser-Induced Breakdown Spectroscopy[J]. Journal of Atmospheric and Environmental Optics, 2019, 14(5): 345

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Apr. 3, 2018

    Accepted: --

    Published Online: Oct. 14, 2019

    The Author Email: YANGLichao (377626867@qq.com)

    DOI:10.3969/j.issn.1673-6141.2019.05.004

    Topics