NUCLEAR TECHNIQUES, Volume. 46, Issue 11, 110402(2023)

Study on the calibration method of newly connected seismic water radon monitor

Rengui HUANG1,2, Ying ZHAO1,2, Yuze LI1,2, Hongtao NING2, Hongyan ZHOU2、*, Hua MAO2, Jian XIAO2, Mengren XIAO1,2, and Pei HUANG3
Author Affiliations
  • 1Observatory for Geodynamic of the East Yangtze Block in Jiujiang, Jiujiang 332006, China
  • 2Jiangxi Earthquake Agency, Nanchang 330039, China
  • 3East China Institute of Technology, Nanchang 330013, China
  • show less

    Background

    The calibration of radon measuring instruments is the most important and key technical link to ensure the reliability and accuracy of radon observation data and plays a vital role in seismic monitoring and prediction.

    Purpose

    This study aims to obtain a calibration method suitable for the calibration of newly connected seismic water radon monitor according to the National Metrological Verification Regulations for Radon Meters (JJG825-2013), the calibration experiment is carried out for the analog radon observatory (manual), which is newly adopted by the seismic network, and the calibration method suitable for the seismic water radon observation is obtained.

    Methods

    Firstly, the FD-125 radon-thorium analyzer was used as the experimental instrument, the calibration experiments were carried out in the standard radon chamber by three calibration methods: circulation method, vacuum method, and flow-gas method. Then, the volume response coefficients of three methods were calculated, and the uncertainty sources and mathematical models were analyzed according to JJG825—2013 regulation. Measured values by internationally recognized standard AlphaGUARDPQ2000Pro radon meter were taken as theoretical references of radon concentration for radon chambers, hence the uncertainties of three calibration methods were obtained.

    Results

    The results show that the circulation calibration method has the highest detection efficiency and fewer influencing factors, and is suitable for manual and digital radon detector calibration. The vacuum calibration method is affected by the standard radon gas diluted by pipeline gas and pressure balance, resulting in low negative pressure sampling efficiency and high volume activity response coefficient R. As for the flow-gas calibration method, the continuous injecting of radon gas into the scintillation chamber affects the dynamic stability of radon gas in the scintillation chamber, and radon gas is not discharged to the background during calibration, resulting in the high volume activity response coefficient R and high intrinsic error.

    Conclusions

    The cyclic method is more suitable for the calibration of new seismic water radon observatories, and this study provides a reference for the calibration of other new water radon observatories and gas radon observatories.

    Tools

    Get Citation

    Copy Citation Text

    Rengui HUANG, Ying ZHAO, Yuze LI, Hongtao NING, Hongyan ZHOU, Hua MAO, Jian XIAO, Mengren XIAO, Pei HUANG. Study on the calibration method of newly connected seismic water radon monitor[J]. NUCLEAR TECHNIQUES, 2023, 46(11): 110402

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Mar. 10, 2023

    Accepted: --

    Published Online: Dec. 23, 2023

    The Author Email:

    DOI:10.11889/j.0253-3219.2023.hjs.46.110402

    Topics