Acta Photonica Sinica, Volume. 46, Issue 3, 304001(2017)

Measurement System for Quantum Efficiency of the Single Photon Detector Based on Correlated Photons

SHI Xue-shun1,2、*, LIU Chang-ming1, ZHAO Kun1, and CHEN Hai-dong1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    The measurement facility based on parametric down-conversion technique was established to measure the quantum efficiency of the near-infrared band single-photon detector. The correlated photons were generated via spontaneous parametric down-conversion in a periodically poled potassium titanyl phosphate crystal pumped by a 518 nm picosecond laser. The detection efficiency of Si-avalanche photodiode and InGaAs-avalanche photodiode single-photon detectors were measured via correlated photon pairs at 778 nm and 1 550 nm, respectively, and the measrement uncertainty was analyzed. The experiment results indicated that the facility can measuer near inferad single-photon detectors quantum efficiency with 1% measrement uncertainty.

    Tools

    Get Citation

    Copy Citation Text

    SHI Xue-shun, LIU Chang-ming, ZHAO Kun, CHEN Hai-dong. Measurement System for Quantum Efficiency of the Single Photon Detector Based on Correlated Photons[J]. Acta Photonica Sinica, 2017, 46(3): 304001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Aug. 15, 2016

    Accepted: --

    Published Online: Apr. 10, 2017

    The Author Email: Xue-shun SHI (xshshi@163.com)

    DOI:10.3788/gzxb20174603.0304001

    Topics