Spectral Calibration for the Large Aperture Infrared Radiometer

YE Zhao1,2、*, WAN Zhi1, LI Xian-sheng1, REN Jian-wei1, and LIU Hong-xing1,2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    The present paper introduced a spectral calibration method to calibrate a large aperture infrared radiation facility. We developed a radiometer which consisted of optical system, infrared detectors(cavity pyroelectric detector and HgCdTe detector 2~14 μm), fine mechanical modulator, lock in amplifier, signal processor, etc. At first, we analyzed the method on how to measure the spectral calibration of the large aperture infrared radiometer, and established the spectral calibration facility. Then, we tested the nonlinear response for the cavity pyroelectric detector and HgCdTe detector. Finally, we used the cavity pyroelectric detector to calibrate the relative spectral responsivity of HgCdTe detector at several wavelengths on the facility. Through the comparison of the two methods for measuring the relative spectral responsivity, the average of multiple measurements and comparative analysis of two methods were given. The uncertainty analysis of the whole system showed that the measurement uncertainty of the facility was better than 3.4%.

    Tools

    Get Citation

    Copy Citation Text

    YE Zhao, WAN Zhi, LI Xian-sheng, REN Jian-wei, LIU Hong-xing. Spectral Calibration for the Large Aperture Infrared Radiometer[J]. Spectroscopy and Spectral Analysis, 2012, 32(7): 1994

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Oct. 3, 2011

    Accepted: --

    Published Online: Sep. 26, 2012

    The Author Email: Zhao YE (sunyeath@gmail.com)

    DOI:10.3964/j.issn.1000-0593(2012)07-1994-05

    Topics