Optoelectronics Letters, Volume. 14, Issue 3, 179(2018)
Time constant optimization ofsolar irradiance absolute radiometer
We experimentally evaluate and optimize the time constant of solar irradiance absolute radiometer (SIAR). The systemic error introduced by variable time constant is studied by a finite element method. The results shown that, with a classic time constant of 30 s for SIAR, the systemic errors are 0.06% in the midday and 0.275% in the morning and afternoon. The uncertainty level which can be considered negligible for SIAR is also investigated, and it is suggested that the uncertainty level has to be less than 0.02%. Then, combining the requirement of international comparison with these two conclusions, we conclude that the suitable time constant for SIAR is 20 s.
Get Citation
Copy Citation Text
TANGXiao, FANG Wei, WANG Yu-peng, YANG Dong-jun, and YI Xiao-long. Time constant optimization ofsolar irradiance absolute radiometer[J]. Optoelectronics Letters, 2018, 14(3): 179
Category: Devices
Received: Nov. 18, 2016
Accepted: Dec. 19, 2016
Published Online: Sep. 15, 2018
The Author Email: Wei FANG (tanggxiao1022@126.com)