Acta Optica Sinica, Volume. 42, Issue 21, 2111001(2022)

Parameter Calibration of Savart Prism Based on Spatially Modulated Polarization Imaging

Ziyang Zhang1,3, Song Ye1,3、*, Wei Xiong2, and Wei Luo1
Author Affiliations
  • 1Guilin University of Electronic Technology, Guilin 541004, Guangxi , China
  • 2Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, Anhui , China
  • 3Guangxi Key Laboratory of Optoelectronic Information Processing, Guilin 541004, Guangxi , China
  • show less

    It is necessary to detect and calibrate the device parameters to reduce errors caused by device parameter deviation during the polarization information extraction of the target. In this paper, the parameters of the Savart prism are calibrated through the beam-splitting imaging experiment on the Savart prism. In addition, a simulation is carried out on the basis of the fractional amplitude polarization camera, and the demodulation effect is quantitatively analyzed by the calculation of the structural similarity and the least square fitting method. Finally, the demodulation diagrams under different polarization states are compared by polarization imaging experiments. The influence of parameter errors on different Stokes components is analyzed, and the effectiveness of the parameter calibration method is verified.

    Tools

    Get Citation

    Copy Citation Text

    Ziyang Zhang, Song Ye, Wei Xiong, Wei Luo. Parameter Calibration of Savart Prism Based on Spatially Modulated Polarization Imaging[J]. Acta Optica Sinica, 2022, 42(21): 2111001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Mar. 10, 2022

    Accepted: May. 23, 2022

    Published Online: Nov. 4, 2022

    The Author Email: Ye Song (yesong@guet.edu.cn)

    DOI:10.3788/AOS202242.2111001