Spectroscopy and Spectral Analysis, Volume. 34, Issue 3, 808(2014)

Determination of Si and P in MnZn Ferrites by Inductively Coupled Plasma Mass Spectrometry

ZHANG Ping1、*, FU Liang2, and NIE Xi-du1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    An inductively coupled plasma mass spectrometry (ICP-MS) was developed for the determination of Si and P in MnZn ferrites. The sample was digested by HNO3+HCl with microwave digestion followed by dilution with ultrapure water then the above two elements in the solution were analyzed directly by ICP-MS. The analytical elements were by introducing the helium gas into the octopole reaction system (ORS) to eliminate the polyatomic interferences caused by the high salty matrixes. The matrix effects and the instrument drift could be commendably calibrated with Y as internal standard element. The working parameters of the instrument were optimized. The NIST SRM 362 and NIST SRM 364 were used as standard reference materials. The results showed a good agreement between the certified values and the experimental results for three analytes. The detection limits for Si and P is 091 and 027 μg·L-1.

    Tools

    Get Citation

    Copy Citation Text

    ZHANG Ping, FU Liang, NIE Xi-du. Determination of Si and P in MnZn Ferrites by Inductively Coupled Plasma Mass Spectrometry[J]. Spectroscopy and Spectral Analysis, 2014, 34(3): 808

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Jun. 3, 2013

    Accepted: --

    Published Online: Mar. 14, 2014

    The Author Email: Ping ZHANG (zphy2008@163.com)

    DOI:10.3964/j.issn.1000-0593(2014)03-0808-04

    Topics