Journal of Applied Optics, Volume. 42, Issue 1, 1(2021)
Analysis on influence factors of MRTD in infrared imaging system
Aiming at the test results of minimum resolvable temperature difference (MRTD) of a multi-band common aperture long focal length reflective infrared imaging system deviating from design value under laboratory conditions, starting with the definition of MRTD, the influences of observation frequency response, visual angle, sampling phase transfer function and other factors affecting MRTD test results were analyzed in detail. Secondly, the influence of sampling phase on infrared system static and dynamic MRTD test results was emphasized. It was demonstrated that static MRTD test value would deviate from the design value when spatial frequency was in the range of
Get Citation
Copy Citation Text
. Analysis on influence factors of MRTD in infrared imaging system[J]. Journal of Applied Optics, 2021, 42(1): 1
Category: OE SYSTEM AND ENGINEERING
Received: Jul. 7, 2020
Accepted: --
Published Online: Apr. 7, 2021
The Author Email: