High Power Laser and Particle Beams, Volume. 33, Issue 8, 086001(2021)
Development of readout electronics system for main detector of CSNS multi-function reflectance spectrometer
A special readout electronic system was developed to meet the needs of the 3He multiwire proportional chamber detector (MWPC) of multi-functional reflectance spectrometer (MR) of China Spallation Neutron Source (CSNS) . The system is mainly composed of core preamplifier board and trigger fan out board. Six preamplifier boards are used to digitize 142 analog signals of the detector. The neutron information is screened through the discrimination mechanism, and the effective neutron cases are packaged and sent to the back end. The trigger fan out board provides the T0 signal and trigger signal arriving at the same time to ensure the alignment of data. The readout electronics system has been tested in laboratory and beam conditions respectively. The test results show that the performance is better than the design requirements. At present, the MWPC detector has been successfully installed in the field of MR spectrometer, and has begun to operate stably.
Get Citation
Copy Citation Text
Lixin Zeng, Yubin Zhao, Shaojia Chen, Xingcheng Tian, Li Yu, Hong Luo, Yanfeng Wang, Dongxu Zhao, Zhijia Sun. Development of readout electronics system for main detector of CSNS multi-function reflectance spectrometer[J]. High Power Laser and Particle Beams, 2021, 33(8): 086001
Category: Nuclear Science and Engineering
Received: Apr. 13, 2021
Accepted: --
Published Online: Sep. 3, 2021
The Author Email: Sun Zhijia (sunzj@ihep.ac.cn)