Acta Photonica Sinica, Volume. 35, Issue 1, 138(2006)

Study of Solution Concentration Measurement System Based on a Method of Glancing Incidence

Jin Qingli*
Author Affiliations
  • [in Chinese]
  • show less

    To essentially eliminate the measuremental errors arisen from light source power and wavelength fluctuation,a measurement method of solution concentration by means of glancing incidence was proposed.The method is based on the relationship between solution concentration and the refraction angle of the medium transparent. Such a measurement system was developed. The relationship between critical light of refraction angle and the concentration was analyzed,introducing the structure of measurement system,and the researching of measured precision and practical. The system to measuring solution concentration of standard,the test data go to regression and demarcate. A mathematical formula is based on the relationship between solution concentration and refraction angle. Different concentrations of a set of sucrose solution,with the concentration being from 5% to 80%,were measured experimentally. The maximum absolute measurement bias is no more than 0.003%. Tests show this system prospectively with practical applications of engineering.

    Tools

    Get Citation

    Copy Citation Text

    Jin Qingli. Study of Solution Concentration Measurement System Based on a Method of Glancing Incidence[J]. Acta Photonica Sinica, 2006, 35(1): 138

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Nov. 12, 2004

    Accepted: --

    Published Online: Jun. 3, 2010

    The Author Email: Qingli Jin (jinqli@wznc.zj.cn)

    DOI:

    CSTR:32186.14.

    Topics