Chinese Optics Letters, Volume. 11, Issue 11, 110401(2013)

Calibration and Monte Carlo simulation of a single-photon counting charge-coupled device for single-shot X-ray spectrum measurements

Yonghong Yan, Lai Wei, Xianlun Wen, Yuchi Wu, Zongqing Zhao, Bo Zhang, Bin Zhu, Wei Hong, Leifeng Cao, Zeen Yao, and Yuqiu Gu

A Princeton Instruments PI-LCX 1300 charge-coupled device (CCD) camera used for X-ray spectrum measurements in laser-plasma experiments is calibrated using three radioactive sources and investigated with the Monte Carlo code Geant4. The exposure level is controlled to make the CCD work in single photon counting mode. A summation algorithm for obtaining accurate X-ray spectra is developed to reconstruct the X-ray spectra, and the results show that the developed algorithm effectively reduces the low-energy tail caused by split pixel events. The obtained CCD energy response shows good linearity. The detection efficiency curves from both Monte Carlo simulations and the manufacturer agree well with the experimental results. This consistency demonstrates that event losses in charge collection processes are negligible when the developed summation algorithm of split pixel events is employed.

Tools

Get Citation

Copy Citation Text

Yonghong Yan, Lai Wei, Xianlun Wen, Yuchi Wu, Zongqing Zhao, Bo Zhang, Bin Zhu, Wei Hong, Leifeng Cao, Zeen Yao, Yuqiu Gu. Calibration and Monte Carlo simulation of a single-photon counting charge-coupled device for single-shot X-ray spectrum measurements[J]. Chinese Optics Letters, 2013, 11(11): 110401

Download Citation

EndNote(RIS)BibTexPlain Text
Save article for my favorites
Paper Information

Category:

Received: Apr. 15, 2013

Accepted: Sep. 25, 2013

Published Online: Nov. 25, 2013

The Author Email:

DOI:10.3788/col201311.110401

Topics