Infrared and Laser Engineering, Volume. 46, Issue 12, 1205002(2017)

Laser source used for measurement absolute spectral responsivity of long-wave infrared detectors

Chen Haidong1、*, Zhao Kun1, Shi Xueshun1,2, Liu Changming1, Liu Yulong1, and Liu Hongbo1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    In order to measure the absolute spectral responsivity of long-wave infrared detectors with high accuracy, improve the long-wave infrared laser power stability and beam quality, a set of long-wave infrared laser source was developed. In this laser source, a feedback-control system, based on the electro-optics effect of CdTe crystal, was designed to improve the stability of long-wave infrared laser power and a long-wave infrared spatial filter, based on the propagation properties of laser beam, was designed to improve the long-wave infrared laser beam quality. The parameters of the spatial filter were also analyzed in this paper. Experimental result shows the laser power stability is elevated to above 0.1% and laser beam quality is improved obviously. The reliable laser source is provided to measure absolute spectral responsivity of long-wave infrared detectors with high-accuracy.

    Tools

    Get Citation

    Copy Citation Text

    Chen Haidong, Zhao Kun, Shi Xueshun, Liu Changming, Liu Yulong, Liu Hongbo. Laser source used for measurement absolute spectral responsivity of long-wave infrared detectors[J]. Infrared and Laser Engineering, 2017, 46(12): 1205002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: 激光器技术

    Received: Apr. 5, 2017

    Accepted: May. 3, 2017

    Published Online: Jan. 19, 2018

    The Author Email: Haidong Chen (haidong4523918@126.com)

    DOI:10.3788/irla201746.1205002