Acta Optica Sinica, Volume. 30, Issue 6, 1680(2010)

High Precision Measurement of Modulation Transfer Function for X-Ray CCD with Knife-Edge Method

Zhang Xiaolin*, Du Guohao, Deng Bao, Xie Honglan, Xue Yanling, Chen Can, Ren Yuqi, Zhou Guangzhao, and Xiao Tiqiao
Author Affiliations
  • [in Chinese]
  • show less

    Knife-edge method is one of the main methods to obtain the modulation transfer function (MTF) of X-ray detectors,while data processing is critical to the accuracy. A series of data processing methods is presented,which can improve the measurement accuracy of MTF for a high-resolution X-ray CCD. Before experiments,MTF simulation is employed to predict the curve trend and noise level. Subsection-analysis algorithm is used to analyze the curve fluctuation in order to eliminate the subjective errors. MTF is recorrected by the oblique angle of the knife edge,which is gotten by using self-adaptive algorithm,clip-window algorithm,and Hough transform. After all the procedures,the mean square deviation of 0.011 for the MTF is achieved,while the Nyquist frequency error is 2.52%.

    Tools

    Get Citation

    Copy Citation Text

    Zhang Xiaolin, Du Guohao, Deng Bao, Xie Honglan, Xue Yanling, Chen Can, Ren Yuqi, Zhou Guangzhao, Xiao Tiqiao. High Precision Measurement of Modulation Transfer Function for X-Ray CCD with Knife-Edge Method[J]. Acta Optica Sinica, 2010, 30(6): 1680

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Jul. 3, 2009

    Accepted: --

    Published Online: Jun. 7, 2010

    The Author Email: Xiaolin Zhang (zhangxiaolin@sinap.ac.cn)

    DOI:10.3788/aos20103006.1680

    Topics