Acta Optica Sinica, Volume. 9, Issue 12, 1119(1989)

Measurement of droplet diameter distribution by laser scattering method

HU ZHUGUO1 and SHENG DEREN2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Theoretical researches and measurement of the mean diameter and the diameter distribution of particles, such as low-concentration water droplets appearing in wet steam turbines, have been done by using an optical measurement apparatus with a He-Ne laser developed by the authors. In the design of experimental system, an anti-Gaussian filter is used to eliminate the effect of Gaussian beam and also a Fourier Transformation lens of high precision is applied to eliminate the spherical aberration In addition, a photosensitive detector is manufactured. By meana of these three deyio^ it is possible to greatly inorease the measurement aoouraoy. This apparatus is oa!ibra1)ep' by directly measuring diameters of glass beads suspended in water. The results are-compared with those obtained by miorophotographio statistical method and with those obtained by using a Malvern instrument. The experimental results indicate that this! light scattering apparatua is oharaoterized by high precision, wide application area andi broad measurement range of particle size (5~200 μm).

    Tools

    Get Citation

    Copy Citation Text

    HU ZHUGUO, SHENG DEREN. Measurement of droplet diameter distribution by laser scattering method[J]. Acta Optica Sinica, 1989, 9(12): 1119

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Sep. 10, 1988

    Accepted: --

    Published Online: Sep. 20, 2011

    The Author Email:

    DOI:

    Topics