Spectroscopy and Spectral Analysis, Volume. 32, Issue 7, 1999(2012)

Analysis of Multi-Micromirrors Tilt Error in Space-Modulated Fourier Transform Infrared Spectrometry

ZHENG Ying1,2、*, LIANG Jing-qiu1, and LIANG Zhong-zhu1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    The system of space-modulated Fourier transform infrared spectrometer without moving parts is stable and compact, moreover, it can realize the on-line monitoring and real-time detection. The position accuracy of multi-micromirrors as the key components has a great influence on spectrometric performance. In the present paper, the multi-mcromirrors tilt error caused by optical path difference changes was calculated and simulated. The analysis shows that, as the multi-micromirrors tilt angle increased, the interference pattern contrast decreased and the signal to noise ratio reduced. The results could be used in the tolerance distribution and the systemic alignment in the next step.

    Tools

    Get Citation

    Copy Citation Text

    ZHENG Ying, LIANG Jing-qiu, LIANG Zhong-zhu. Analysis of Multi-Micromirrors Tilt Error in Space-Modulated Fourier Transform Infrared Spectrometry[J]. Spectroscopy and Spectral Analysis, 2012, 32(7): 1999

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Nov. 13, 2011

    Accepted: --

    Published Online: Sep. 26, 2012

    The Author Email: Ying ZHENG (opticszhy@163.com)