Acta Optica Sinica, Volume. 35, Issue 8, 812005(2015)

Computer-Aided Alignment Technology of Multi-Spectral Camera

Zhang Chao*, Wei Xuemin, Zhao Xiting, Jiao Wenchun, and Liao Zhibo
Author Affiliations
  • [in Chinese]
  • show less

    Refract optical system using in remote sensing usually has many elements,complicated structure and diffraction limited image quality,which makes the alignment puzzled.Combined with the adjustment mathematical model,the CODE V optical design software and the field interference Zernike coefficient are integrated into the alignment variables for calculating the misalignment.According to the off-axis three mirror ideal optical system model,the alignment sensitivity matrix is acquired and the correct order of multi misalignment variables is determined,which can overcome the alignment correlation effectively.It is effective and convenient to raise the optical system wavefront error root mean square (RMS) from 0.563 λ (=632.8 nm) to 0.064 λ .After the orbit test,the multi-spectral camera imaging quality is excellent.

    Tools

    Get Citation

    Copy Citation Text

    Zhang Chao, Wei Xuemin, Zhao Xiting, Jiao Wenchun, Liao Zhibo. Computer-Aided Alignment Technology of Multi-Spectral Camera[J]. Acta Optica Sinica, 2015, 35(8): 812005

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Apr. 8, 2015

    Accepted: --

    Published Online: Jul. 29, 2015

    The Author Email: Chao Zhang (zhangchao19870408@163.com)

    DOI:10.3788/aos201535.0812005

    Topics