Chinese Optics Letters, Volume. 2, Issue 3, 03174(2004)

Study on the wavelength calibration of type III concave grating spectrometry system

Li Bai*, Ningfang Liao, Zhaojian Li, and Weiping Yang
Author Affiliations
  • National Laboratory of Color Science and Engineering, School of Information Science and Technology, Beijing Institute of Technology, Beijing 100081
  • show less

    We discuss and calibrate the spectrometry system based on concave reflection grating. The working principle, structure and parameters of the spectrometry system are introduced. For the wavelength calibration problem, three methods are put forward and discussed in detail with formulation calculation method, circular iteration method and interpolation. Interpolation is used to calibrate the concave reflection grating spectrometry system and the error is less than 1 nm. Four spectrum images that the system collected are given in this paper. The experimental results indicate that a spectrometry system can be based on concave reflection grating and be calibrated by interpolation.

    Tools

    Get Citation

    Copy Citation Text

    Li Bai, Ningfang Liao, Zhaojian Li, Weiping Yang. Study on the wavelength calibration of type III concave grating spectrometry system[J]. Chinese Optics Letters, 2004, 2(3): 03174

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Nov. 11, 2003

    Accepted: --

    Published Online: Jun. 6, 2006

    The Author Email: Li Bai (shurimemphis@163.com)

    DOI:

    Topics