Infrared and Laser Engineering, Volume. 51, Issue 9, 20210984(2022)

Measurement technique research for the absorptivity of cryogenic radiometer absorbing cavity at the 4 K temperature

Bing Yu1,2, Jihong Fan2, Linguang Yuan2, Yan Li2, Lei Guo2, Xiao Wang3, Junwei Chu2, Yan Qin2, Yunan Sun2, Deng Zhang2, Yue You2, and Weiqi Jin1、*
Author Affiliations
  • 1MOE Key Laboratory of Optoelectronic Imaging Technology and System, Beijing Institute of Technology, Beijing 100081, China
  • 2Xi′an Institute of Applied Optics, Xi′an 710065, China
  • 3Chongqing Institute of Green Intelligent Technology, Chinese Academy of Sciences, Chongqing 710699, China
  • show less

    In order to realize measuring the absorptivity of cryogenic radiometer absorbing cavity at the 4 K temperature, a method of measuring the absorptivity with variable temperature is researched. By designing reflection monitoring components in front of the Brewster window of the cryogenic radiometer, and controlling the cryogenic radiometer to work in a vacuum environment of 10-6 Pa, then adjusting the refrigeration temperature of the cryogenic radiometer, the reflection signals of the cryogenic radiometer absorbing cavity are measured at 632.8 nm under room temperature and different temperature conditions. Combined with the measurement results of the reflectance at 632.8 nm of the cryogenic radiometer absorbing cavity at room temperature using the traditional integrating sphere method, the absorptivity of the cryogenic radiometer absorption cavity can accurately obtain under different temperature conditions through calculations. The experimental measuring the absorptivity of the absorbing cavity at room temperature and 4 K temperature, the absorptivity is 0.99976 and 0.999 71, respectively. The measurement uncertainty of the absorptivity of the cryogenic radiometer absorption cavity at the 4 K condition is evaluated, and the results obtained show that the relative expanded uncertainty is 0.005%(k=2).

    Tools

    Get Citation

    Copy Citation Text

    Bing Yu, Jihong Fan, Linguang Yuan, Yan Li, Lei Guo, Xiao Wang, Junwei Chu, Yan Qin, Yunan Sun, Deng Zhang, Yue You, Weiqi Jin. Measurement technique research for the absorptivity of cryogenic radiometer absorbing cavity at the 4 K temperature[J]. Infrared and Laser Engineering, 2022, 51(9): 20210984

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Photoelectric measurement

    Received: Dec. 17, 2021

    Accepted: Jan. 13, 2022

    Published Online: Jan. 6, 2023

    The Author Email:

    DOI:10.3788/IRLA20210984