Acta Photonica Sinica, Volume. 41, Issue 9, 1059(2012)

A Calibration Technique of Sensitivity of Fiber Grating Hydrophone

CHEN Yi1、*, ZHANG Jun1, JIN Xiaofeng2, and HUANG Shengye2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    In order to calibrate the sensitivity of fiber grating hydrophone, a calibration technique of fiber grating hydrophone was studied. In this paper, the working principle of fiber grating hydrophone was introduced firstly. Secondly, using tunable laser and intensity modulation method with workpoint controlling, the signal demodulation and stably measurement of fiber grating hydrophone was realized. Based on these methods, a calibration system of fiber grating hydrophone was established using comparison methods, and its calibration frequency range was 20 Hz to 10 kHz. Finally, using this calibration system, the sensitivity of a fiber grating hydrophone made of equivalent phase shift fiber Bragg grating was calibrated in the frequency range 20 Hz to 10 kHz, and measurement uncertainty of the system was estimated. The results proved that this system can correctly calibrate the sensitivity of fiber grating hydrophone, with its expand uncertainty (at k=2) of 0.9 dB.

    Tools

    Get Citation

    Copy Citation Text

    CHEN Yi, ZHANG Jun, JIN Xiaofeng, HUANG Shengye. A Calibration Technique of Sensitivity of Fiber Grating Hydrophone[J]. Acta Photonica Sinica, 2012, 41(9): 1059

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Mar. 29, 2012

    Accepted: --

    Published Online: Aug. 31, 2012

    The Author Email: Yi CHEN (y.chen@163.com)

    DOI:10.3788/gzxb20124109.1059

    Topics