Laser & Optoelectronics Progress, Volume. 60, Issue 12, 1217002(2023)

Main Factor Analysis for Imaging Performance of Hand-Held Full-Field Optical Coherence Tomography System with Dual Interferometers

Ran Li, Wanrong Gao*, and Yifeng Tang
Author Affiliations
  • School of Electronic and Optical Engineering, Nanjing University of Science & Technology, Nanjing 210094, Jiangsu, China
  • show less

    In this study, we report an improved hand-held full-field optical coherence tomography (FFOCT) system based on a combination of double interferometers. The system uses a Fizeau interferometer for detection and a Michelson interferometer for compensation. The initial system was modified by the addition of a plane mirror to alter its optical path and to make it easier to arrange samples horizontally. The analysis was conducted on how the system's imaging quality and compensation distance were affected by the angle deviation between the mirror and other key components of the system. The obtained results are useful for the design and optimization of the hand-held FFOCT system.

    Tools

    Get Citation

    Copy Citation Text

    Ran Li, Wanrong Gao, Yifeng Tang. Main Factor Analysis for Imaging Performance of Hand-Held Full-Field Optical Coherence Tomography System with Dual Interferometers[J]. Laser & Optoelectronics Progress, 2023, 60(12): 1217002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Medical Optics and Biotechnology

    Received: Jun. 5, 2022

    Accepted: Jun. 20, 2022

    Published Online: Jun. 1, 2023

    The Author Email: Gao Wanrong (wgao@njust.edu.cn)

    DOI:10.3788/LOP221765

    Topics