Acta Optica Sinica, Volume. 33, Issue 8, 830001(2013)
Experimental Study of the Photochemical Reaction in the Smog Chamber Using a Chernin Multipass Cell
A Chernin multipass absorption spectroscopy system is combined with the photochemical simulation smog chamber and applied to the detection of trace gases in the smog chamber. The system accuracy and reliability are proved with NO2 detection. The NO2 measurements of the system and the results of Thermo 42i NO-NO2-NOx analyzer have a good linear relationship. The system is applied to online detection of SO2 during β-pinene/O3/SO2 system chemical reaction in the smog chamber and with 36 m absorption path length, the SO2 detection limit of the system is 22.2×10-9. The result demonstrates that the Chernin multipass absorption spectroscopy system has the capability of quantitative measurement and can be used in online detection of trace gases during photochemical reactions in the smog chamber.
Get Citation
Copy Citation Text
Cheng Yue, Zhao Weixiong, Hu Changjin, Gu Xuejun, Pei Shixin, Huang Wei, Zhang Weijun. Experimental Study of the Photochemical Reaction in the Smog Chamber Using a Chernin Multipass Cell[J]. Acta Optica Sinica, 2013, 33(8): 830001
Category: Spectroscopy
Received: Feb. 25, 2013
Accepted: --
Published Online: Jul. 9, 2013
The Author Email: Yue Cheng (chengyue@aiofm.ac.cn)