Spectroscopy and Spectral Analysis, Volume. 35, Issue 4, 971(2015)

Trace Carbon Monoxide Detection with a Cavity Ring-Down Spectrometer

CHEN Bing1、*, ZHOU Ze-yi2, KANG Peng1, LIU An-wen1, and HU Shui-ming1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    A cavity ring-down spectrometer (CRDS) was built based on telecom distributed feedback (DFB) diode lasers.The ring down cavity was formed by mirrors of 99.997% reflectivity with a separation of 130 cm,with an empty ring down time of about 150 μs.A minimum detectable absorption coefficient of 5×10-12 cm-1 was obtained by averaging about 1 000 recorded spectra.The ring down cavity is thermo-isolated,and used as an interferometer to calibrate the recorded spectrum.A feedback control scheme was applied to step scan the laser frequency,successively matching each of the longitudinal modes of the cavity.By measuring the CO contents in a standard gas sample,the quantitative capability of the CRDS instrument was demonstrated,and a CO detection limit of 4 ppb was achieved.The instrument was applied to monitor the CO concentration in ambient air.

    Tools

    Get Citation

    Copy Citation Text

    CHEN Bing, ZHOU Ze-yi, KANG Peng, LIU An-wen, HU Shui-ming. Trace Carbon Monoxide Detection with a Cavity Ring-Down Spectrometer[J]. Spectroscopy and Spectral Analysis, 2015, 35(4): 971

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Mar. 3, 2014

    Accepted: --

    Published Online: Apr. 20, 2015

    The Author Email: Bing CHEN (bingchen@mail.ustc.edu.cn)

    DOI:10.3964/j.issn.1000-0593(2015)04-0971-04