Chinese Optics Letters, Volume. 9, Issue 10, 101201(2011)

Precise force measurement method by a Y-shaped cavity dual-frequency laser

Guangzong Xiao, Xingwu Long, Bin Zhang, and Geng Li
Author Affiliations
  • Department of Optoelectronic Engineering, College of Optoelectronic Science and Engineering, National University of Defense Technology, Changsha 410073, China
  • show less

    A novel precise force measurement based on a Y-shaped cavity dual-frequency laser is proposed. The principle of force measurement with this method is analyzed, and the analytic relation expression between the input force and the change in the output beat frequency is derived. Experiments using a 632.8-nm Y-shaped cavity He-Ne dual-frequency laser are then performed; they demonstrate that the force measurement is proportional to a high degree over almost five decades of input signal range. The maximum scale factor is observed as 5.02 \times 10^9 Hz/N, with beat frequency instability equivalent resolution of 10^{-5} N. By optimizing the optical and geometrical parameters of the laser sensor, a force measurement resolution of 10^{-6} N could be expected.

    Tools

    Get Citation

    Copy Citation Text

    Guangzong Xiao, Xingwu Long, Bin Zhang, Geng Li. Precise force measurement method by a Y-shaped cavity dual-frequency laser[J]. Chinese Optics Letters, 2011, 9(10): 101201

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, measurement, and metrology

    Received: Jan. 18, 2011

    Accepted: May. 9, 2011

    Published Online: Jul. 13, 2011

    The Author Email:

    DOI:10.3788/col201109.101201

    Topics